Cảm biến đo gió siêu âm 3-D Campbell Scientific CSAT3B

Thương hiệu: Campbell Scientific | Tình trạng: Còn hàng
Liên hệ

Cảm biến đo gió siêu âm 3-D Campbell Scientific CSAT3B

Gọi 0904251826 để được trợ giúp

Thiết bị đo gió siêu âm 3-D Campbell Scientific CSAT3B

Overview

The CSAT3B is an update and replacement to the original CSAT3.

The CSAT3B has an aerodynamic design, operates in a pulsed acoustic mode, withstands exposure to harsh weather conditions, and is well suited for eddy-covariance measurements. Because the electronics are packaged inside the mounting block of the CSAT3B head, installation is easier and the placement of the instrument is flexible.

Benefits and Features

Integrated electronics that provide easy mounting of a single piece of hardware

Integrated inclinometer

High-precision measurements ideal for turbulence and eddy-covariance studies

An improved design with a thin, aerodynamic support strut close to the ends of the sensor arms, creating greater rigidity and improved accuracy of sonic temperature

Datalogger sampling supported for any frequency between 1 and 100 Hz

Automatic status check for faults or window contamination

New CPI communications for more robust, higher bandwidth measurements

Multiple communication options including SDM, CPI, USB, and RS-485

Internal temperature and humidity measurements with easily replaced desiccant

Version 5 algorithm for calculating data outputs; combines the signal sensitivity of version 3 with the rain performance of version 4

Includes options to filter high frequencies for applications requiring analysis of non-aliased spectra

Specifications

Operation Temperature Range -30 to +50°C (equivalent to 312 to 368 m s-1 in speed of sound)
Outputs ux, uy, uz, Ts (ux, uy, uz are wind components referenced to the anemometer axes; Ts is sonic temperature in degrees Celsius.)
Speed of Sound Determined from three acoustic paths. (Corrected for crosswind effects.)
Wind Direction Range 2.5 to 357.5° in CSAT3B coordinate system (0 to 360° customized)
Filter Bandwidths 5, 10, 20, or 25 Hz
Measurement Path Length 10.0 cm (3.9 in.) vertical; 5.8 cm (2.3 in.) horizontal
Transducer Angle from Horizontal 60 degrees
Transducer Diameter 0.64 cm (0.25 in.)
Transducer Mounting Arm Diameter 0.84 cm (0.33 in.)
Support Arm Diameter 1.59 cm (0.63 in.)
Anemometer Head Weight 1.45 kg (3.2 lb)
Anemometer Dimensions 60.64 x 12.2 x 43.0 cm (23.87 x 4.8 x 16.9 in.)

Wind Accuracy

-NOTE- Accuracy specifications assume the following:
  • -30° to +50°C operating range
  • Wind speeds < 30 m s-1
  • Wind angles between ±170°
Maximum Offset Error < ±8.0 cm s-1 (ux, uy), < ±4.0 cm s-1 (uz)
Maximum Gain Error
  • < ±2% of reading (wind vector within ±5° of horizontal)
  • < ±3% of reading (wind vector within ±10° of horizontal)
  • < ±6% of reading (wind vector within ±20° of horizontal)

Measurement Resolution

ux, uy 1 mm s-1 rms
uz 0.5 mm s-1 rms
Ts ± 0.002°C RMS (at 25°C)
Wind Direction < 0.058° (ux = uy ≤ 1 m s-1)

Measurement Rates

Datalogger Triggered 1 to 100 Hz
Unprompted Output (to PC) 10, 20, 50, or 100 Hz
Internal Self-Trigger Rate 100 Hz

Measurement Delay

Datalogger Triggered (no filter) 1 trigger period (1 scan interval)
Unprompted Output (no filter) 10 ms
Filtered Output (datalogger-prompted or unprompted to PC)
  • 795 ms (with 5 Hz bandwidth filter)
  • 395 ms (with 10 Hz bandwidth filter)
  • 195 ms (with 20 Hz bandwidth filter)
  • 155 ms (with 25 Hz bandwidth filter)

Internal Monitor Measurements

Update Rate 2 Hz
Inclinometer Accuracy ± 1°
Relative Humidity Accuracy
  • ± 3% (over 10 to 90% range)
  • ± 7% (over 0 to 10% range)
  • ± 7% (over 90 to 100 % range)
Board Temperature Accuracy ± 2°C

SDM

-NOTE- Used for datalogger-based data acquisition.
Bit Period 10 µs to 1 ms
Cable Length
  • 7.6 m (25 ft) max (@ 10 µs bit period)
  • 76 m (250 ft) max (@ 1 ms bit period)
Address Range 1 to 14
Bus Clocks per Sample ~200

CPI

-NOTE- Used for datalogger-based data acquisition.
Baud Rate 50 kbps to 1 Mbps
Cable Length
  • 15 m (50 ft) max (@ 1 Mbps)
  • 122 m (400 ft) max (@ 250 kbps)
  • 853 m (2800 ft) max (@ 50 kbps)
Address Range 1 to 120
Bus Clocks per Sample ~300

RS-485

-NOTE- Used for configuration or PC-based data acquisition.
Baud Rate 9.6 kbps to 115.2 kbps
Cable Length
  • 305 m (1000 ft) max (@ 115.2 kbps)
  • 610 m (2000 ft) max (@ 9.6 kbps)
Bus Clocks per Sample ~500 (ASCII formatted)

USB

-NOTE- Used for configuration or PC-based data acquisition.
Connection Speed USB 2.0 full speed 12 Mbps
Cable Length 5 m (16.4 ft) maximum

Power Requirements

Voltage Supply 9.5 to 32 Vdc
Current at 10 Hz Measurement Rate
  • 110 mA (@ 12 Vdc)
  • 65 mA (@ 24 Vdc)
Current at 100 Hz Measurement Rate
  • 145 mA (@ 12 Vdc)
  • 80 mA (@ 24 Vdc)

Download

Brochures

CSAT3B 3D Sonic Anemometer

Manuals

CSAT3B Three-Dimensional Sonic Anemometer

CSAT3B Packing Information

Miscellaneous

CSAT3B Cabling Options Diagram

 